Assistance 4D probe; realize the probe via the ID code; swap the 4D push signal by means of the  electrical relay to produce the 4D probe work Usually. 4.two.two Principal board The main board might be divided into two components: entrance-end of the leading board and again-end of the principle board.

Maximum four moments tasking for a person transmitted beam, leading to fantastic time resolution and higher frame price.

Get a complete and extended watch on the anatomical structure as a result of panoramic imaging coupled with velocity indicator and ahead/backward scan ability making scanning easier, smoother plus more controllable.

Do not use any cables in addition to the cables furnished With all the device by Mindray. 3. Use the cable provided with This technique to attach the printer. Other cables may result in electric shock. Page fifteen: Mechanical Protection

Minimize off cable ties connecting the leading device over the Screen by diagonal chopping pliers, after which you can pull out the corresponding cable plugs.

Mindray’s new technology of black & white ultrasound system, DP-50 Pro, is supplied with Highly developed imaging systems allowing for deeper penetration and faster graphic acquisition with larger resolution.

Web site one hundred forty five Cleanse control panel Applications: dry delicate fabric, gentle soapy water  Approach:  Use dry delicate cloth to scrub the floor in the system. In case the program is dirty, moisten the delicate fabric with delicate soapy drinking water and wipe off any stains. Use dry delicate fabric to remove any dampness and permit all really hard surfaces to absolutely air-dry. Webpage 146: Thoroughly clean The Peripherals

Explore greater diagnostic information via prolonged perspective with the anatomical composition on all convex probes.

Analyzing clients in today’s demanding atmosphere demands innovative technological innovation that permits clinicians to produce fast and confident diagnoses. The flexibility on the DP-fifty expands your clinical utility while assisting to boost affected person care.

5.two.three Geometric positioning precision Longitudinal geometric positioning accuracy  Take a look at Step: Do changes as the best way in tests the most depth. Report the gap by 20mm each segment over the longitudinal targets line utilizing the measurement caliper; Decide on the worth with the best mistake (to 20mm), estimate the precision using the components under The measurement caliper really should be positioned at the higher edge of the focus on, Notice:... Website page 79 Transverse geometric positioning accuracy  Exam Step: Address the scan surface of your phantom with h2o or few gel, Carefully Get hold of the probe With all the scan surface area Adjust the depth, making the transverse targets for being shown from the impression. Modify the main focus point to be posited beside the transverse targets (the conventional is just not obvious) Modify parameters like gain, TGC, making Each and every transverse targets to generally be Plainly displayed. Page 80 5.five.2.four Blackout Location Take a look at Stage: Go over the scan surface area in the phantom with h2o or few gel, Carefully Call the probe While using the scan surface area Adjust the depth in a decreased benefit, and set the main focus at the nearest destination to the scan surface area. Minimize the value of parameters like AP, Gain right up until the history sounds just might be witnessed. Web site 81: Computer software Update And Upkeep

Kebutuhan Anda akan alat profesional dan canggih yang dikombinasikan dengan pencitraan berkualitas dan kemudahan kegunaan kini dapat dipenuhi dengan sistem portabel baru Mindray DP-50Exp.

Be aware: Just before uploading the log, be sure the ultrasound technique has been linked to community, if not, the technique may alert “could not connect with server” The server (smtp.

General performance Take a look at five.5.1 Take a look at System Resolution Detection depth Geometric Situation Accuracy Blackout area test History the Examination images five.5.two Exam Information The image utilised Here's just for reference, keep on with the image effect in the real NOTE: problem. Specifications: Show: established the distinction and brightness in the medical software benefit (or even the default position) Procedure atmosphere: dim area, simulating the clinical application atmosphere. Webpage 76 In issue the transverse resolution tests targets are horizontally shown, history the minimum distance of Dp50rb login two targets which can be clearly acknowledged. Repeat the Procedure higher than to the transverse resolution testing targets at other depths. As demonstrated in determine beneath. Axial resolution ... Page 77 When utilizing the convex probe, hold the transverse resolution testing targets to NOTE: be displayed close to the midline. When employing a linear probe with steer purpose, never turn on the steer functionality when conduct the transverse resolution exam. Zoom within the area wherever the targets Situated if required. The diameter of your goal place at a particular depth is equivalent to the transverse resolution within the depth. Website page seventy eight 5.

Immediately after changing the CPU module, key file needs to be created Again according to the new MAC deal with and created into HDD of the device. Or else, the configuration functionality on the item might be dropped. Right after getting rid of the CPU module, Clear away panhead screws with washers (12 M3X6) preset on the principle device box, you will take from the board. Page 118 Figure eight-29 Disassemble the User interface Assembly (1) Maintain the higher edge of control panel assembly, and open up the user interface. Figure 8-thirty Disassemble the Control Panel Assembly (2) Right after removing 1 countersunk head self-tapping screw PT3X10 preset on the user interface and consider out the plug from the keyboard and the main unit connecting cables, you might get rid of the control panel assembly.

A smart new shape, enhanced mobility and easy Procedure make DP-fifty well suited for all clinical settings.